Titel: |
Titel:
Calibration of climatic chambers : Guideline DKD-R 5-7
|
Autoren: |
Autoren:
Deschermeier, Regina, Physikalisch-Technische Bundesanstalt (PTB), Fachbereich 3.4, Analytische Chemie der GasphaseDillmann, Eugen, Weiss Technik GmbH, Balingen Friederici, Sven, Physikalisch-Technische Bundesanstalt (PTB), Fachbereich 7.4, Temperatur Gutknecht, Harald, Thermo Electron LED GmbH, Langenselbold Hager, Helmut, Techmetrics GmbH, Winnenden Alle Autoren anzeigen (16) |
Beitragende: |
Beitragende:
HostingInstitution: Physikalisch-Technische Bundesanstalt (PTB), ISNI: 0000 0001 2186 1887
|
Seiten: |
Seiten:
68
|
Sprachen: |
Sprachen:
en
|
DOI: |
DOI:
10.7795/550.20250403
|
Version: |
Version:
01/2025
|
Art der Ressource: |
Art der Ressource:
Standard / Technical Requirement
|
Verlag: |
Verlag:
Physikalisch-Technische Bundesanstalt (PTB)
|
Rechte: |
Rechte:
https://creativecommons.org/licenses/by-nc-nd/3.0/de/CC by-nc-nd 3.0 |
Beziehungen: |
Beziehungen:
IsVariantFormOf: DOI 10.7795/550.20250120
|
Datumsangaben: |
Datumsangaben:
Verfügbar: 2025-04-07 Erstellt: 2025-01 |
Datei: |
Datei:
Datei herunterladen
(application/pdf)
1.13 MB (1185171 Bytes)
MD5 Prüfsumme: 1e0d78f06855ddb2d7127230abfbc524 SHA256 Prüfsumme: a0b15c7af1a8bbdedf77a2c1ac854eb6d4c4af83e417d7859a2093e887df8768 |
Stichwörter: |
Stichwörter:
DKD guideline ;
calibration ;
climatic chamber ;
rel. humidity ;
temperature ;
DIN; 13277:2022 ;
DIN 12880:2007 ;
DIN EN 60068
|
Zusammenfassung: |
Zusammenfassung:
This guideline lays down the minimum requirements for the calibration procedure and the determination of the measurement uncertainty in the calibration of climatic chambers. It is applicable to the calibration of climatic chambers for gas temperature and relative gas humidity or only for gas temperature. In the context of this guideline, gas humidity is understood to mean gaseous material systems consisting exclusively of the components water and carrier gas. The scope of the guideline covers the carrier gases air and nitrogen. For other carrier gases proof of validation is required. This guideline also applies to the calibration of individual measuring locations in climatic chambers; in this case, the calibration is not valid for the entire climatic chamber.
|
Zitat: |
Zitat:
Guideline DKD-R 5-7, Calibration of climatic chambers, Edition 01/2025, Revision 0, Physikalisch-Technische Bundesanstalt, Braunschweig and Berlin. DOI: 10.7795/550.20250403
|