Application of the substitution procedure for the calibration of non- automatic weighing instruments : Guideline DKD-R 7-3
Autoren:
Autoren:
Osang, Steffen, Minebea Intec Bovenden, Bovenden
Häfner, Martin, Häfner Gewichte GmbH, Oberrot
Fritsch, Klaus, Mettler-Toledo (Schweiz) GmbH, Greifensee, SWITZERLAND
Müller, Klaus, Mettler-Toledo GmbH, Gießen
Ortmann, Jochen, Hessische Eichdirektion, Darmstadt

Alle Autoren anzeigen (7)
Beitragende:
Beitragende:
HostingInstitution: Physikalisch-Technische Bundesanstalt (PTB), ISNI: 0000 0001 2186 1887
Seiten:
Seiten:
14
Sprachen:
Sprachen:
en
DOI:
DOI:
10.7795/550.20250306
Version:
Version:
01/2025
Art der Ressource:
Art der Ressource:
Standard / Technical Requirement
Verlag:
Verlag:
Physikalisch-Technische Bundesanstalt (PTB)
Rechte:
Rechte:
https://creativecommons.org/licenses/by-nc-nd/3.0/de/
CC by-nc-nd 3.0
Beziehungen:
Beziehungen:
IsVariantFormOf: DOI 10.7795/550.20250129
Datumsangaben:
Datumsangaben:
Verfügbar: 2025-03-07
Erstellt: 2025-01
Datei:
Datei:
Datei herunterladen (application/pdf) 736.36 kB (754031 Bytes)
MD5 Prüfsumme: 1d7cd460e1eceefc02bc0c82d5219402
SHA256 Prüfsumme: 1dfb4046addca9bb6ec97db6e1e199d472727876e7f941d1af15dc5ee629326c
Stichwörter:
Stichwörter:
DKD Guideline ; substitution procedure ; DKD-R 7-2 ; EURAMET Calibration Guide No. 18 chapter 4.3.3 ; calibration ; non-automatic weighing instrument
Zusammenfassung:
Zusammenfassung:
In many cases there are not enough reference weights available for calibrations of non-automatic electronic weighing instruments. Moreover, placing the reference weights on the weighing instrument may prove to be difficult if the load receptor lacks sufficient space. Therefore, the substitution procedure is an appropriate option for calibrating high-capacity scales. The guideline contains additional requirements regarding the practical implementation of the substitution procedure, thus supporting the high competence required of an accredited calibration laboratory. The process of the substitution procedure is clearly illustrated by means of a specific example. The guideline helps to standardise the substitution procedure in the calibration of non-automatic weighing instruments. It should be noted that the measurement uncertainty during calibration and during use is significantly increased by the substitution procedure. The requirements regarding the use of the substitution procedure aim to ensure the practicality and comparability of the calculation of the uncertainty of measurement during calibration and when using the weighing instrument. The guideline is based on the calibration guideline DKD-R 7-2, Edition 01/2018 translation of the EURAMET Calibration Guide No.18 Version 4.0 (11/2015)].
Zitat:
Zitat:
Guideline DKD-R 7-3, Application of the substitution procedure for the calibration of non-automatic weighing instruments, Edition 01/2025, Revision 0, Physikalisch-Technische Bundesanstalt, Braunschweig and Berlin. DOI: 10.7795/550.20250306.

Autoren

Osang, Steffen, Minebea Intec Bovenden, Bovenden
Häfner, Martin, Häfner Gewichte GmbH, Oberrot
Fritsch, Klaus, Mettler-Toledo (Schweiz) GmbH, Greifensee, SWITZERLAND
Müller, Klaus, Mettler-Toledo GmbH, Gießen
Ortmann, Jochen, Hessische Eichdirektion, Darmstadt
Grünewälder, Carsten, C. Grünewälder Waagentechnik, Wuppertal
Mecke, Jonas, Physikalisch-Technische Bundesanstalt (PTB), Fachbereich 1.1, Masse

PTB-OAR Menü

Sprache wechseln: uk flag