Measurement uncertainty budget for torque moment at measurement flexure during the operation of the torque standard machine
Expected Value M_T_{5 MNm} = 114.229 Nm

Measurement uncertainty budget
Quantity	Unit	Value	Standard uncertainty	Distribution	Sensitivity coefficient	Uncertainty contribution	Index
X_i		x_i	u_i		c_i	c_i*u_i	(c_i*u_i/u_c)^2
k_gamma_MF	N/mm	0.381E6	3240	Normal	0.00030	0.97	99.1 %
\epsilon_TSM	rad	80.3E-6	23.2E-6	Rectangular	-110	-0.0026	0.0 %
k_gamma	-	0.957970	779E-6	Rectangular	120	0.093	0.9 %
u_RR	mm	0.4668	46.7E-9	Normal	240	0.057E-3	0.0 %
r_RR	mm	1493.75	0.067	Normal	-0.076	-0.0051	0.0 %

#M_T	= torque moment applied on the measurement flexure during the calibration in the 5 MNm torque standard machine
#k_gamma_MF	= M_T stiffness coefficient
#\epsilon_TSM	= misalignment of the measurement flexure
#k_gamma	= coefficient to calculate the displacement at the specific measurement flexure via the rotation angle, found in the finite element analysis
#u_RR	= displacement of the retroreflector, measuring the measurement lever rotation in the torque standard machine
#r_RR	= distance from the pivot point of the measurement lever to the retroreflector position